ELISA Washer Throughput: From Cycle Time to Comparable Batch Capacity

ELISA washer throughput is useful only when the measurement boundary and wash protocol are clear. A specification for dispensing liquid, a timed wash program and the number of acceptable plates completed in a working batch describe different things. Compare like-for-like conditions before using any of them to plan a process.

This guide provides a calculation method and a blank trial record. It does not report measured performance for a machine or promise a production rate.

01 / Performance measurement

Start with the protocol, not the headline rate

Record the plate format, occupied wells, wash cycles, volume per well, aspiration settings, soak duration and shaking steps. Include any initial aspiration and final aspiration. Keep the protocol version with the results so that a faster run cannot be mistaken for the same method after settings change.

Manufacturer specifications illustrate why conditions matter. Agilent’s 405 TS technical data sheet associates wash-speed figures with plate format, liquid volume and cycle count, and separately lists programmable soaking. Those published conditions describe that manufacturer’s equipment. They are not performance evidence for another washer or for your complete batch.

A dispensing-only time excludes liquid removal unless the test explicitly includes it. A single wash cycle also differs from a program containing several wash cycles and timed soaks. Never convert one into a complete-process claim without accounting for the remaining steps.

02 / Performance measurement

Define three measurement boundaries

Program time: Start when the wash program begins and stop when it finishes. Include every programmed wash, soak and aspiration step. State whether plate positioning is included.

Batch elapsed time: Start at the agreed preparation boundary and stop when the last plate completes the defined batch operation. Include loading, unloading, priming, buffer changes, refills, waste handling and interruptions within that boundary. State whether final cleaning is included; for repeatable batch comparisons, include it or show it separately.

Complete-process time: Follow the plates through all required operations, including incubation and transfers. A wash-stage result cannot be described as finished-kit output. Define the endpoint precisely, since coating, drying, packaging and release may lie beyond the washer’s scope.

03 / Performance measurement

Calculate rates from recorded time

For a program measured in seconds:

Program-only equivalent plates/hour = 3,600 ÷ measured seconds per plate.

Label this as a program-only equivalent. It assumes uninterrupted repetition and excludes anything outside that timer. It is not an achieved batch rate.

For a completed batch:

Accepted plates/hour = accepted unique plates × 60 ÷ batch elapsed minutes.

Count each accepted plate once. Record rejected plates and rework separately; rework consumes time but does not create an additional unique plate. Define acceptance before the trial, using the relevant residual-liquid, consistency and process-quality criteria. If acceptance depends on later testing, mark the count provisional until those results are available.

Also record operator minutes. Two systems with similar accepted output may require different amounts of hands-on work. Do not substitute operator time for elapsed time in the throughput denominator.

04 / Performance measurement

Include overhead without double counting

For a strictly sequential process, a planning estimate can add setup and priming, repeated plate handling and program time, buffer-change time, cleaning and other stops. Keep each event in one category. Do not add a soak separately if it is already included in the program timer.

For overlapping operations, use the actual batch timeline. Loading, washing and incubation may happen simultaneously, so adding their individual durations can overstate elapsed time. Conversely, ignoring a queue can overstate capacity. Record start and finish timestamps rather than assuming ideal overlap.

Incubation deserves its own capacity check. Berthold’s coating guidance includes incubation between liquid-handling steps. Record the required duration, available plate positions and transfer windows. Even if incubation occurs away from the washer, insufficient space can delay the next batch. Count on-washer dwell time as equipment occupancy.

05 / Performance measurement

Blank batch-throughput record

Use the same template for each candidate configuration and protocol. Repeat representative runs and retain the individual results rather than reporting only the fastest.

Blank worksheet. Scroll within the table on a narrow screen.

Record field Trial entry
Date, operator, machine and configuration ____
Plate type, wells used and batch identifier ____
Protocol/version; cycles; volume per well ____
Aspiration; soak; shaking settings ____
Start/finish boundary and timestamps ____
Program time and plates timed ____
Loading/unloading time and method ____
Priming and buffer-change time/volume ____
Refills, waste handling, cleaning ____
Stops: reason, duration, affected plates ____
Incubation duration, positions, queue time ____
Unique processed/accepted/rejected plates ____
Rework time and acceptance record ____
Total elapsed minutes; operator minutes ____
Accepted plates/hour; exclusions ____

06 / Performance measurement

Frequently asked questions

Can I compare two advertised plates-per-hour figures directly?

Only after matching protocol, configuration and timing boundaries. Otherwise, treat them as incomplete inputs for a trial.

Should downtime be removed?

Keep it in the batch result when it occurred within the stated boundary. You may report program-only performance separately, clearly labeled.

What if there are no measured records?

Leave the output blank. Use the template to collect evidence and label any planning calculation as an estimate with stated assumptions.

Discuss your requirements

Share the process you need to evaluate

Tell us your plate format, required operations, liquids and batch schedule. Include the evidence and acceptance criteria your team needs to assess a configuration.

Use the form, our contact page, or info@elisa-washer.com.

Tell us about your project

Include what you know and identify anything that still needs review.

    Read our Privacy Policy before sending enquiry details.